      內窺鏡維修常見的故障有什么?

      來源:http://www.hhmhwz.com/ 發布時間:2025-07-11 瀏覽量:0

        隨著醫療技術發展,醫用內窺鏡在臨床廣泛應用,使用中常出現光源、氣腹、液體加壓等系統故障,掌握相應解決辦法可保障設備正常運行。

        With the development of medical technology, medical endoscopes are widely used in clinical practice. System failures such as light source, pneumoperitoneum, and liquid pressurization often occur during use. Mastering corresponding solutions can ensure the normal operation of the equipment.

        光源系統故障中,光線發暗多因燈泡接近使用壽命(氙光燈泡壽命約 500 小時),排除光柵遮擋后更換新燈泡即可,帶壽命計時裝置的設備更易判斷。開機燈泡不亮若非燈泡燒毀,多為高壓點火電路元件老化,注意不可用萬用表檢測脈沖高壓。

        In the case of a malfunction in the light source system, dim light is often caused by the bulb approaching its service life (xenon light bulbs have a lifespan of about 500 hours). After eliminating grating obstruction, replacing the bulb with a new one is sufficient. Equipment with a lifespan timer is easier to determine. If the light bulb does not light up when turned on, it is mostly due to the aging of high-voltage ignition circuit components, and it is important not to use a multimeter to detect pulse high voltage.

        人工氣腹系統故障需針對性處理:壓力顯示異常多是氣體壓力傳感器損壞,若為多組壓差傳感器,建議整體更換避免同組元件陸續失效;流量不準更換流量或壓差傳感器;設備漏氣常因密封圈老化、減壓閥或電磁閥損壞,較易判斷修復。

        The malfunction of the artificial pneumoperitoneum system needs targeted treatment: abnormal pressure display is often caused by damaged gas pressure sensors. If it is multiple sets of differential pressure sensors, it is recommended to replace them as a whole to avoid the failure of components in the same set one after another; Do not replace flow or differential pressure sensors for flow; Equipment leakage is often caused by aging sealing rings, damage to pressure reducing valves or solenoid valves, which are easier to diagnose and repair.

        液體加壓系統常見蠕動泵損壞、液體壓力傳感器故障(原理同氣體傳感器)及內部電路損壞,需對應更換部件或檢修電路。

        Common damage to peristaltic pumps, liquid pressure sensors (based on the same principle as gas sensors), and internal circuits in liquid pressurization systems require corresponding replacement of components or maintenance of circuits.

      20200403031114827

        電凝電切系統故障中,腳踏開關、負極板導線等配件易損,直接更換即可;電切、電凝及雙極故障,可根據現象維修或更換對應電路模塊。

        In the case of faults in the electrocoagulation and cutting system, accessories such as foot switches and negative electrode wires are easily damaged and can be replaced directly; Electric cutting, electrocoagulation, and bipolar faults can be repaired or replaced with corresponding circuit modules according to the symptoms.

        動力系統主要是刨削刀頭磨損和手柄內電機問題,電機更換受型號限制,需注重使用保護;消融系統多為刀頭損耗,設備故障常為電路問題,可從輸出部分排查。

        The main issues with the power system are the wear of the planing blade and the motor inside the handle. The replacement of the motor is limited by the model, and attention should be paid to its protection during use; The ablation system is mostly caused by blade wear, and equipment failures are often circuit problems, which can be investigated from the output part.

        沖洗吸引系統壓力或吸力不足,可能是雙向泵老化或負壓瓶漏氣;設備無法開機若因液體誤吸,清理干凈未損壞部件即可恢復。

        Insufficient pressure or suction in the flushing suction system may be due to aging of the bidirectional pump or leakage of the negative pressure bottle; If the device cannot be turned on due to liquid aspiration, it can be restored by cleaning the undamaged parts.

        為減少故障,建議設專用手術間,由專科護士操作保養,嚴格遵循術前檢查、術中排障、術后維護流程。設備故障優先采取應急維修,可解決約 85% 的中小故障。

        To reduce malfunctions, it is recommended to set up a dedicated operating room, operated and maintained by specialized nurses, and strictly follow the procedures of preoperative examination, intraoperative troubleshooting, and postoperative maintenance. Priority should be given to emergency maintenance for equipment failures, which can solve about 85% of small and medium-sized faults.

        本文由內窺鏡設備維修友情奉獻.更多有關的知識請點擊:http://www.hhmhwz.com我們將會對您提出的疑問進行詳細的解答,歡迎您登錄網站留言.

        This article is a friendly contribution from the maintenance of endoscopic equipment For more information, please click: http://www.hhmhwz.com We will provide detailed answers to your questions. You are welcome to log in to our website and leave a message

      s
      新聞資訊
      聯系我們
      • 公司公司:匠仁醫療設備有限公司
      • 匠仁醫療山東濟南維修部
      • 聯系人:樊經理
      • 電話:13153199508
      • 地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室
      • 匠仁醫療湖南長沙維修部
      • 聯系人:李經理
      • 電話:13873135765
      • 地址:湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室
      新聞資訊

      樊經理:13153199508     李經理:13873135765

      公司地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室  湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室

           

      備案號:魯ICP備2023027194號-1    魯公網安備 37010402441281號

      主站蜘蛛池模板: 国产免费人成视频尤勿视频 | 亚洲精品成a人在线观看☆| 亚洲日韩在线视频| 亚洲人成毛片线播放| 亚洲AV电影天堂男人的天堂| 免费看黄福利app导航看一下黄色录像| ww在线观视频免费观看w| **一级一级毛片免费观看| 成人免费无码大片a毛片| 青青草原亚洲视频| 亚洲欧洲精品久久| 亚洲国产成人无码AV在线影院| 免费无码A片一区二三区| 伊人久久综在合线亚洲91| 三年片在线观看免费西瓜视频| 午夜国产精品免费观看| 亚洲天堂中文字幕在线| 亚洲国产精品久久人人爱| 久久久久久99av无码免费网站| 日本亚洲欧美色视频在线播放| 亚洲av午夜成人片精品电影| 亚洲人成电影在线观看青青| 免费视频淫片aa毛片| 亚洲av成人无码久久精品| 男人扒开添女人下部免费视频| 国产综合精品久久亚洲| 99视频有精品视频免费观看| 国产亚洲av片在线观看18女人| 男男gvh肉在线观看免费| 在线v片免费观看视频| 久久精品国产亚洲av高清漫画 | 久久久久亚洲av无码尤物| 国产在线a免费观看| 高潮毛片无遮挡高清免费视频| 最近中文字幕无吗免费高清 | 国产v片免费播放| 亚洲中文字幕无码mv| 18女人毛片水真多免费| 亚洲欧洲免费无码| 国产免费卡一卡三卡乱码| 国产拍拍拍无码视频免费|